Image striping when acquiring data with a commutator - discussion #35
Comments
Which commutator are you using? e.g. the Open Ephys commutator is a mil-spec grade RF rotary joint that maintaines 50-ohm transmission line through the commutation point (less chance for data corruption than slip ring with no impedance control) that we have testing for hundreds of hours continuous use without issue. https://open-ephys.org/commutators/coaxial-commutator and that is compatilbe with miniscope v4. |
Currently using a passive slip ring commutator from Moog that uses a 50Ohm coaxial line for the scope. |
OK, yes. That does not maintain 50-ohm transmission line. Introduces discontinuity which results in signal distortion. This is exacerbated by the fact that its right in the middle of the cable (assuming you have a sizable length of coax on either side). This may not be the issue, but it certainly wont help. |
Yeah I agree the occasional discontinuity from the slip ring is the root cause. The OEPS commutator only support 1 scope for purchase (right?), and I'm using two miniscopes through the commutator and keeping the cabling in and out fairly short (1.5-2.0m on animal side, 5-10cm on DAQ side). I see the OEPS design for 2 scopes is available on the dev site, but I don't have the capacity to fabricate one right now. I was hoping there could be a software solution to detecting and resolving the striping. Perhaps using a real-time template matching or just a user interface to reset the buffer if they notice it occurring to prevent the stable offset situation from persisting too long during a recording. |
I developed a decent fix for the static banding issue - see the demo video and matlab script in my miniscope banding repo edit - @daharoni fyi, if this helps things |
This is intended to be more of a discussion and spot to organize and document ideas, as the google discussion is a bit dated and long.
Image striping during acquisition
During acquisition occasionally the image becomes misaligned due to a buffer issue. Seems to occur more often when using a commutator. I'm aware this issue has be documented and discussed elsewhere (see google groups discussion, I just want to create a more centralized issue to track progress and solutions.
I've been working a decent amount on identifying and fixing them, but haven't come up with a satisfying method. Currently I manually find the beginning and end of each occurrence, then interpolate them out using the bookend frames (so that motion correction and source extraction don't go haywire), and then throw out the activity from those interpolated times. Usually this only results in a handful of lost frames at a time, but when it lasts hundreds of frames it's very detrimental.
I've found it helpful to classify the stripes into two categories - stable and moving (see examples below).
Here's a link to example data showing both issues:
https://drive.google.com/drive/folders/1JoomTZ7Ty0XKJo9ZNjIwy6_CUKqlzK_N?usp=sharing
Screenshots
All of the following occurred in sequence:
Normal frame
Example offset pattern
Another example offset
Example of moving (non-stable) striping 1
Example of moving (non-stable) striping 2
Back to normal imaging (after about 135 frames)