You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
‘Run profile > stop’ will stop the experiment profile wherever its last leaving off (Ex. consider a profile that starts stirring at 400 RPM and stops after 1.5 mins. If stirring is active, and the profile is stopped early, it will keep stirring indefinitely) — does this make sense for users? Would it be better to stop all jobs associated with the profile?
Starting a new profile with stirring and temp automation, while temp is already running. The profile successfully starts but the temp is skipped, and a yellow notif appears after the green success notification. The profile will successfully finish and not indicate that any actions couldn't be accomplished. Suggestion = to not allow a profile to begin unless all profile activities can succeed, or allow “starting” a profile to restart/reconfigure any active activities.
Experiment profile activities can be stopped via pioreactor > manage > stop on individual activities, but the profile will still appear to be running and will still “successfully execute”. However, when clicking “stop all activities”, the profile will “successfully” exit early and state ‘actions not completed’. These two “stop” actions should have the same result.
The text was updated successfully, but these errors were encountered:
Bug = stirring is still running even after experiment profile has completed + success notification. Activities in the pioreactor menu show that stirring is not running. Starting+stopping stirring, and stopping all activities will not stop this stirring...
Bug = stirring is still running even after experiment profile has completed + success notification. Activities in the pioreactor menu show that stirring is not running. Starting+stopping stirring, and stopping all activities will not stop this stirring...
Can you provide detailed replication steps for this one?
Some issues/suggestions:
‘Run profile > stop’ will stop the experiment profile wherever its last leaving off (Ex. consider a profile that starts stirring at 400 RPM and stops after 1.5 mins. If stirring is active, and the profile is stopped early, it will keep stirring indefinitely) — does this make sense for users? Would it be better to stop all jobs associated with the profile?
Starting a new profile with stirring and temp automation, while temp is already running. The profile successfully starts but the temp is skipped, and a yellow notif appears after the green success notification. The profile will successfully finish and not indicate that any actions couldn't be accomplished. Suggestion = to not allow a profile to begin unless all profile activities can succeed, or allow “starting” a profile to restart/reconfigure any active activities.
Experiment profile activities can be stopped via pioreactor > manage > stop on individual activities, but the profile will still appear to be running and will still “successfully execute”. However, when clicking “stop all activities”, the profile will “successfully” exit early and state ‘actions not completed’. These two “stop” actions should have the same result.
The text was updated successfully, but these errors were encountered: