If the PIFOC has a range of 100 microns but the user asks for 300 microns of stack, then the acquisition will proceed with no errors but most of the data will be redundant.
The text was updated successfully, but these errors were encountered:
raacampbell
changed the title
There is no check that the Z-drive has sufficient range to image the requested range of depths
There is no check that the PIFOC has sufficient range to image the requested range of depths
May 21, 2024
This is a bit tricky because if the user has altered the position with the offset knob, then it won't be clear where the PIFOC is. Should be possible to at least return an error if the stack exceed the maximum capacity of the PIFOC. If the feedback of the PIFOC indicates it's not at zero then this could also serve as a warning that we might be getting things wrong.
If the PIFOC has a range of 100 microns but the user asks for 300 microns of stack, then the acquisition will proceed with no errors but most of the data will be redundant.