Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Buzzer review #44

Open
Benhalor opened this issue Oct 9, 2020 · 0 comments
Open

Buzzer review #44

Benhalor opened this issue Oct 9, 2020 · 0 comments

Comments

@Benhalor
Copy link
Collaborator

Benhalor commented Oct 9, 2020

The fundamental frequency of the high-priority signal must be between 150 and 1000Hz (it is now equal to 4075Hz).

Pulse duration is not respected (High priority alarm must consist of 5 pulses repeated twice (the current state has 2))

image
image

@Benhalor Benhalor changed the title Alarm review Buzzer review Oct 17, 2020
Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment