Check ms/us overflows #48

Open
pi-r-p opened this issue Oct 9, 2020 · 0 comments
Open

Check ms/us overflows #48

pi-r-p opened this issue Oct 9, 2020 · 0 comments

Comments

@pi-r-p
Copy link
Collaborator

pi-r-p commented Oct 9, 2020

Each time millis or micros is called in the code, overflow must be fixed.
There is one 71 minutes overflow, and one 50 days overflow.
A medical device may stay powered during 50 days.