Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

restart position without the MiKNOW UI #35

Open
zchatt opened this issue Feb 24, 2022 · 5 comments
Open

restart position without the MiKNOW UI #35

zchatt opened this issue Feb 24, 2022 · 5 comments

Comments

@zchatt
Copy link

zchatt commented Feb 24, 2022

Hello ONT,

We are trying to run our minion directly from the command line. For this we are using the start_protocol.py.
We initially had an authentication issue (Issues #34), however we set "--no-tls True" as a workaround.
Now when running we receive the following error:

`No flow cell present in position MN39033 (running)`

However, the flow cell is definitely inserted in the minion. I can see that "position_connection.device.get_flow_cell_info()" is used to obtain the flow cell information which is defined in minknow_api.device_service.DeviceService.
Notably, using "position_connection = spec.position.connect()" I can see that the device is running:

	[FlowCellPosition('localhost', {name: "MN39033"
	state: STATE_RUNNING
	rpc_ports {
	  secure: 8001
	  insecure: 8000
	  secure_grpc_web: 8002
	}
	})]

Also, using "position_connection.device.get_device_info()" I can return information on the device;

	device_info device_id: "MN39033"
	max_channel_count: 512
	max_wells_per_channel: 4
	digitisation: 8192

So cool, I can see the device, just not the flow cell. If restart the position using the MiKNOW UI the start_protocol.py script progresses (alas to another issue);

	Failed to find protocol for position MN39033 (running)
	Requested protocol:
	  product-code: None
	  kit: SQK-LSK109
	  basecalling: False
	  basecall_config: None
	  barcode-kits: None
	  barcoding: False

However, I can now retrieve the flow cell info using "position_connection.device.get_flow_cell_info()" after restarting the position with MiKNOW UI

          has_flow_cell: true
          channel_count: 126
          wells_per_channel: 1
          has_adapter: true
          adapter_id: "FA-05846"
          temperature_offset: 327.6700134277344
          asic_version: "IA02D"
          asic_id_str: "7735955"

So this leads to my question, how do we restart the position without the MiKNOW UI? I am probably missing something obvious and I apologize in advance. Thank you for any help.

@zchatt zchatt changed the title restart position restart position without the MiKNOW UI Feb 24, 2022
@0x55555555
Copy link
Contributor

Hello,

https://github.com/nanoporetech/minknow_api/blob/master/proto/minknow_api/manager.proto#L54

reset_position will restart a position, from the manager service.

You shouldn't need to do this though... it is odd your device reports not having a flowcell - how does it display in the UI at the same time? Are you giving the device enough time to initialise - waiting for a flow cell to be detected?

Thanks,

  • George

@zchatt
Copy link
Author

zchatt commented Feb 28, 2022

Thank you for your response. I have left the device for mins to hours but the message does not change.
The MiKNOW UI displays "No flow cell detected" and this does not change not until I right click and "Restart position" . I have tried to reset the position using

manager.reset_position('MN39033')

however this does not return a response and does not result in a flow cell being detected as I am still receiving the error.

"No flow cell present in position MN39033 (running)"

I am using the latest MinION Release 21.11.8. Notably, using an older version (19.12.5) detects the flow cell immediately.

@0x55555555
Copy link
Contributor

It sounds like there may be issues with minknow outside of the API, if the UI also doesnt show a flowcell?

It may be best to get in contact with support, supply logs for the box you are running minknow on, and a description of the problem, and we can work to solve that issue first.

Hope that helps,

  • George

@zchatt
Copy link
Author

zchatt commented Mar 1, 2022

So I have tried this on another machine (Ubuntu 20.04.3) and again the UI does not detect a flow cell unless I "Restart position". Once the flow cell is detected in the UI, the minknow_api can then detect a flow cell eg. "position_connection.device.get_flow_cell_info()"

However, I cannot reproduce "Restart position" of the UI to detect the flow cell using manager.reset_position(). Would this indicate the issue is with minknow_api ?

Happy to log a job with support. Very much appreciate your help.

@0x55555555
Copy link
Contributor

However, I cannot reproduce "Restart position" of the UI to detect the flow cell using manager.reset_position(). Would this indicate the issue is with minknow_api ?

No, still if the position does not display a flow call shortly after one is connected it seems like a more general hardware or software problem - best to contact support so we can sort that out first.

Thanks,

  • George