KCRM datacubes appear to have a 180 degree rotation relative to KCWI datacubes. The below screenshot shows KCRM (left) and KCWI (right) cube slices from observations of a standard star with locked WCS.
From a brief investigation the KCWI datacube has the correct WCS (other than #1674) and it is the KCRM data that is rotated.
The text was updated successfully, but these errors were encountered:
Thanks @ddobie - As discussed on the Users Slack, this is because KCRM has an extra mirror (so the image is flipped relative to KCWI). This issue is resolved in the branch kcrm_Sep23, and will be merged into the development branch shortly after a few more quality controls. You are one of the first to be reducing KCRM data!