[feature] : ASD - Add support for 9.99 volt images #42

Closed
SylviaWhittle opened this issue May 2, 2024 · 0 comments · Fixed by #50
Closed

[feature] : ASD - Add support for 9.99 volt images #42

SylviaWhittle opened this issue May 2, 2024 · 0 comments · Fixed by #50
Assignees
Labels
enhancement New feature or request

Comments

@SylviaWhittle
Copy link
Collaborator

Is your feature request related to a problem?

A collaborator sent over a file that cannot be loaded using our .asd reader, as its analogy-digital range hex value is 0x00000003 which has not yet been added to the list of known values. Strangely, when experimenting with this file, the voltage range is 0 to 9.99 volts (precisely this value, no more, no less else the image height scales break).

Describe the solution you would like.

Add an option in the asd loader for the 0x00000003 hex code and corresponding unipolar voltage range 0 - 9.99V. I have MVP code, just needs cleaning up.

Describe the alternatives you have considered.

N/A

Sample Image

I do not have permission to share the image, but I have a working fix, so I just need to formalise it.

Additional context