HLX503: post-experiment tweaks #7168
Comments
support: ISISComputingGroup/EPICS-hlx503#7 |
We are going to test on the real sorb tomorrow afternoon so will make changes if needed. |
branches patched over to WISH, seems to be working fine and ran overnight with some tests which looked OK according to instrument scientists. Still some issues with a first recondense being temperamental but not introduced by this ticket. To review check db files look OK. IOC tests don't exist on master just yet as I need to sort them out as part of #6240 |
PS drop outs were caused by incorrect default stop bit setting which was different to the ITC503 - have modified to match the ITC and that has solved the comms issues! |
As an instrument scientist using the ITC powered 3he fridge for the first time (this was on WISH) I would like some changes to be made.
Acceptance Criteria
What is the acceptance criteria?
Following issues are fixed: