Skip to content

Commit

Permalink
GMP : Empty Timestamp, Instrument not reset -- abort the GMP run at t…
Browse files Browse the repository at this point in the history
…he end of LIV_UPDATE, inform user of violation of the GMP compliance...
  • Loading branch information
alexsav815 committed Dec 7, 2024
1 parent 6d96d88 commit 009eb26
Show file tree
Hide file tree
Showing 4 changed files with 469 additions and 12 deletions.
87 changes: 78 additions & 9 deletions programs/us_xpn_viewer/us_xpn_viewer_gui.cpp
Original file line number Diff line number Diff line change
Expand Up @@ -2660,8 +2660,30 @@ void US_XpnDataViewer::check_for_sysdata( void )
if ( !timer_data_reload->isActive() )
{
qDebug() << "TRY PROCEED INTO == 5/0 from check_for_sys_data()....";

export_auc_auto();

bool tmstampOK = true;
export_auc_auto( tmstampOK );
if ( !tmstampOK )
{
in_reload_check_sysdata = false;

QMessageBox::critical( this,
tr( "GMP Compliance Problem:" ),
tr( "This run did not start at zero RPM because the w2t value "
"was non-zero at the beginning of the experiment. \n\n"
"Therefore, the boundary conditions are unknown and the experiment "
"cannot be reliably analyzed. It fails GMP requirements and will "
"therefore be aborted. The run will be deleted.\n\n"
"The data can be retrieved with the \"View Raw Optima Data\" function "
"in the UltraScan Utilities." ) );


delete_autoflow_record(); /** Do we want to delete GMP run NOW? **/
reset_auto();
emit aborted_back_to_initAutoflow();

return;
}

// QString mtitle_complete = tr( "Complete!" );
// QString message_done = tr( "Experiment was completed. Optima data saved..." );
Expand Down Expand Up @@ -2974,7 +2996,8 @@ void US_XpnDataViewer::end_processes( void )

QString msg_sys_text = QString(tr("Attention! UltraScan GMP is not able to communicate with the data acquisition server on the %1.\n\n "))
.arg(xpndesc);
QString msg_sys_text_info = QString(tr("The program will <b>Return</b> to \"Manage Optima Runs\" where you can re-attach to this run later "
QString msg_sys_text_info = QString(tr("The program will <b>Return</b> to \"Manage Optima Runs\" "
"where you can re-attach to this run later "
"by clicking \"Select Optima Run to follow\" once the network "
"issue is resolved. UltraScan will then resume data acquisition.\n\n"
"NOTE: If the network connection cannot be re-established to the ongoing run, "
Expand Down Expand Up @@ -3667,7 +3690,30 @@ DbgLv(1) << "RDa: allData size" << allData.size();

if ( !timer_check_sysdata->isActive() ) // Check if sys_data Timer is stopped
{
export_auc_auto();
bool tmstampOK = true;
export_auc_auto( tmstampOK );
if ( !tmstampOK )
{
in_reload_all_data = false;

QMessageBox::critical( this,
tr( "GMP Compliance Problem:" ),
tr( "This run did not start at zero RPM because the w2t value "
"was non-zero at the beginning of the experiment. \n\n"
"Therefore, the boundary conditions are unknown and the experiment "
"cannot be reliably analyzed. It fails GMP requirements and will "
"therefore be aborted. The run will be deleted.\n\n"
"The data can be retrieved with the \"View Raw Optima Data\" function "
"in the UltraScan Utilities." ) );


delete_autoflow_record(); /** Do we want to delete GMP run NOW? **/
reset_auto();
emit aborted_back_to_initAutoflow();

return;
}

updateautoflow_record_atLiveUpdate();

reset_auto();
Expand Down Expand Up @@ -4775,7 +4821,7 @@ void US_XpnDataViewer::connect_ranges( bool conn )
}

// export_auc data in us_com_project -- CORRECTED
void US_XpnDataViewer::export_auc_auto()
void US_XpnDataViewer::export_auc_auto( bool& tmstampOK )
{
inExport = true;

Expand All @@ -4787,7 +4833,7 @@ void US_XpnDataViewer::export_auc_auto()
if ( noptsy == 1 )
{
correct_radii(); // Perform chromatic aberration radius corrections
nfiles = xpn_data->export_auc( allData );
nfiles = xpn_data->export_auc_auto( allData, tmstampOK );
}

//--- Combined optics type -----//
Expand All @@ -4810,7 +4856,7 @@ void US_XpnDataViewer::export_auc_auto()

cb_optsys->setCurrentIndex( osx );
correct_radii(); // Chromatic aberration correction if needed
int kfiles = xpn_data->export_auc( allData ) - 2; // Export data
int kfiles = xpn_data->export_auc_auto( allData, tmstampOK ) - 2; // Export data
nfiles += kfiles; // Total files written
}

Expand Down Expand Up @@ -5332,8 +5378,31 @@ DbgLv(1) << "RLd: NO CHANGE";
{

qDebug() << "Exporing/Writing to disk...";
// ALEXEY Export AUC data: devise export_auc_auto() function which would return directory name with saved data - to pass to emit signal below...
export_auc_auto();
// ALEXEY Export AUC data: devise export_auc_auto() function which would return directory name with saved data - to pass to emit signal below...
bool tmstampOK = true;
export_auc_auto( tmstampOK );
qDebug() << "tmstampOK? " << tmstampOK;
if ( !tmstampOK )
{
in_reload_auto = false;

QMessageBox::critical( this,
tr( "GMP Compliance Problem:" ),
tr( "This run did not start at zero RPM because the w2t value "
"was non-zero at the beginning of the experiment. \n\n"
"Therefore, the boundary conditions are unknown and the experiment "
"cannot be reliably analyzed. It fails GMP requirements and will "
"therefore be aborted. The run will be deleted.\n\n"
"The data can be retrieved with the \"View Raw Optima Data\" function "
"in the UltraScan Utilities." ) );


delete_autoflow_record(); /** Do we want to delete GMP run NOW? **/
reset_auto();
emit aborted_back_to_initAutoflow();

return;
}

// QString mtitle_complete = tr( "Complete!" );
// QString message_done = tr( "Experiment was completed. Optima data saved..." );
Expand Down
2 changes: 1 addition & 1 deletion programs/us_xpn_viewer/us_xpn_viewer_gui.h
Original file line number Diff line number Diff line change
Expand Up @@ -378,7 +378,7 @@ class US_XpnDataViewer : public US_Widgets
void runDetails ( void );

void export_auc ( void );
void export_auc_auto( void );
void export_auc_auto( bool& );

void changeRadius ( void );
void changeCellCh ( void );
Expand Down
Loading

0 comments on commit 009eb26