Skip to content

We use a pair of Keithley 6221 and 2182A to measure Pulse Signals

License

Notifications You must be signed in to change notification settings

lordonium/Pulse-Measurement-System

Repository files navigation

                          LabVIEW Dependencies Contributions welcome License

Basic Overview

While conducting an experiment you can use different techniques for measuring the signals. I was asked to code Keithley 2182A Nanovoltemeter in conjunction with 6221 DC and AC Current Source to perfom Pulse Delta and Delta measurements.

Theory

Before explaining the software part lets tackle a little bit of theory.

Delta

Delta voltage measurements use a current-reversal technique to cancel the effects of thermal EMFs in the voltmeter test lead connections. The simplest algorithm for a delta measurement uses the following 2-point source/measure process:

  1. Source a positive current through the DUT and measure the voltage (V1).
  2. Source a negative current through the DUT and measure the voltage (V2).
  3. Calculate the delta reading as follows:

equation

Authors

  • Sergey Chernayev - Initial work - lordonium

All other known bugs and fixes can be sent to [email protected]