Skip to content

Setting Up Overview

Rob Campbell edited this page Apr 5, 2019 · 19 revisions

Before you start

  • You will need a suitable PC and a functioning install of ScanImage 5. Testing has been done with the latest release of ScanImage 5.5 but BakingTray should also work with its premium equivalent and likely also with later versions.
  • Linear scanners will work, but a resonant scanner is recommended for speed and because the system is better tested in resonant mode. Keep in mind that higher speed Cambridge resonant scanners have a smaller angular range and will produce images with more noticeable shot noise since the pixel dwell times are shorter. An 8 kHz scanner is recommended.
  • You will need a 3-axis stage built from supported motion hardware. You may use different hardware, as it straightforward to write the control classes, but V-508 or V-551 stages from PI are strongly recommended.
  • The stages should be aligned with the imaging plane.
  • If you wish to cut the sample after each tile-scan you will need to bolt a vibrotome next to the objective. The procedure for doing this with a Leica unit is described by Economo et al. Also see Ragan et al for details on sample preparation.

Installing Drivers

  • Install MATLAB. BakingTray was tested on 2016a through 2018a and should work on later releases.
  • If using PI stages, install the PI drivers for the stages and follow the instructions for setting up the MATLAB PI interface. If you need to install the SDK in order to compile the PI drivers, you may find this link useful. The PI MATLAB code will need to be in your path. It should be located at C:\Users\Public\PI\PI_MATLAB_Driver_GCS2 once installed.

Software installation

This proceeds in two steps: 1) Installation and set up of ScanImage. 2) Installation and set up of BakingTray.

Installation and configuration of ScanImage

Install and set up BakingTray

Clone this wiki locally