Notifications
You must be signed in to change notification settings - Fork 5
Upgrading ScanImage
Rob Campbell edited this page Jun 7, 2019
·
4 revisions
- Quit ScanImage
- Unpack ScanImage into a new directory. For example, your current install may be at
C:\MATLAB\SI_R50
and you unpack the new version inC:\MATLAB\SI_R51
- Copy your machine data file into the new directory. Usually you can just use the same file.
- Rename your old install directory to something else (e.g.
C:\MATLAB\SI_R50_OLD
). - Make a backup copy of this old install somewhere else (e.g. a remote server) in case you need to revert to it.
- Add the new ScanImage directory to your MATLAB path. Remove the old one. Start ScanImage.
If you are using USR and CFG files to store customisations, you will add these now. I assume you are keeping these in your ScanImage directories.
- Copy all of the to the config files from your last install to the new directory.
- Load your
.usr
file (ignore any errors) then press save. - Load the
.cfg
you normally use then press save (ignore any errors). Then save the.usr
file so the.cfg
will be loaded along with it. - Repeat loading and saving for any other files or file combinations you have.
- Close ScanImage and re-open it, selecting a
.usr
file to load at startup if applicable. ScanImage should now parse the config files without errors
If you are using a resonant scanner you likely have used the hSI.hScan2D.plotFov2VoltageCurve
tool to ensure you are getting square images.
This information will not have been ported over by the above steps.
To port it over, do the following:
- Close ScanImage
- Copy the
ConfigData\YourScannerName_ResScanner_classData.mat
file from the old install to the new install. - Re-start scanimage. There should be no errors, if there are please file a BakingTray support issue.
- Run
hSI.hScan2D.plotFov2VoltageCurve
and you should see your old voltage curve.
Installation: Getting Started
Hardware requirements
Setting up: Overview
Verifying hardware operation
Starting BakingTray
Setting up ScanImage
Settings Files
Achieving high stitching accuracy
Installation: Calibration
Basic calibrating procedures
Calibrating the stages
Fine-tuning positioning accuracy
Further User Instructions
FAQ
Problems & Solutions