CAEN
Wiki > The Backend System > Specific Device IOC > Power Supplies > CAEN
A high voltage power supply normally used for detector electronics.
There are now two IOCs devoted to the CAEN. An older one from the Canadian light source (HVCAEN) and a newer one from SLAC (HVCAENA). They are communicated with over Ethernet. HVCAEN is used widely and HVCAENA is to be used on DETMON to test it. HVCAENA has more information available including board and crate parameters. Both IOCs have a read-only mode which can be set via macros.
A single IOC can communicate with up to 8 CAEN crates, this appears to be a limit in the CAEN library as it defines MAX_CRATES as 8 in the library header and has a "too many connections" error code listed in its potential error codes.
To do this the crates must be added into the st.cmd
as so:
CAENx527ConfigureCreate "hv0", "_IP_ADDR1_"
CAENx527ConfigureCreate "hv1", "_IP_ADDR2_"
When CAENx527DbLoadRecords
is called it will then create PVs of the form hv0:_SLOT_:_CHANNEL_
The pv setup is a little weird for this device, it looks like value is the setpoint and value:fbk
is the readback. So:
pwonoff
: sets the power
pwonoff:fbk
: readback the value (I am unsure whether this is the setpoint or the status)
The HVCAEN SIM IOC uses a simulated caen library to simulate the CAEN. This does not work completely, current it looks like setting anything does not updates the monitor values.
The SIM also disconnects and reconnects itself every X commands to show that is working.
Finally it looks like the macros SIM
, DEVSIM
and RECSIM
are not respected so you need to set the SIM PV directly.
If you run dbl
and get mainly standard support IOC PVs check that comms have been established to the given IP address
If you log in to the console and cannot get an EPICS prompt, the whole console/procserv must be restarted, as typically the Ctrl-X
option does not work in this situation either
If you can ping and telnet to the crate, but the EPICS driver cannot connect, the crate may need a (physical) power cycle. This is best done by detector group.
You can set HVCAENx527Debug X
in the console (or via the st.cmd) with X being a debug level e.g. 5 and the IOC will print debugging information based on the level you have given.
Made up of three levels the IOC level (st.cmd, config etc.), the CPP/Db support level and the CAENHVWrapper level. The CAENHVWrapper
is a vendor library that contains methods to set and get data to/from CAEN crates, it is documented in the pdf files that can be found in the CAENHVAsyn
support module. The CPP support level was created by SLAC and has been edited by us. The IOC level was created by us to enable use in IBEX.
Currently can communicate with 4 CAEN crates at a time (but very extensible to add more). Set up by defining macros IP_ADDRESS_X and SYS_TYPE_X for X from 0 to 3. Another macro READ_ONLY defines whether we can write to the CAEN crate with 1 for read-only and 0 for read/write.
A crate is loaded in the st.cmd with calls to the support level:
$(IFIP0_PRESENT) CAENHVAsynSetEpicsPrefix("$(MYPVPREFIX)$(IOCNAME):HV0:")
$(IFIP0_PRESENT) CAENHVAsynConfig("HV0",$(SYS_TYPE_0),"$(IP_ADDRESS_0)","username","password")
CAENHVAsynSetEpicsPrefix
simply sets the pv prefix in the driver.
CAENHVAsynConfig
creates a new CAENHVAsyn
object, which loads dbs and creates pvs for system properties, board parameters and channel parameters.
Read/write is set in the st.cmd for all crates with a call to the support level:
CAENHVAsynReadOnly($(READ_ONLY=0))
Set READ_ONLY to 1 for readonly and 0 for read/write.
See UML below for diagram description.
At initialisation of the CAENHVAsyn
object, the crates, boards, channels and their properties and parameters are retrieved from the CAENHVWrapper
and the support objects are populated from this. The process of retrieving and populating objects is kicked off by the creation of the Crate object (call ICrate:create
).
From this data, the dbs are loaded. There are templates for records e.g. ai, ao. These templates are loaded CAENHVAsyn
for board parameters, channel parameters and system properties, with fields filled out by passing macros. The read Rd
records are loaded with a scan rate of 1 second and write/set St
records are set to scan passive. The loading of dbs can be seen in the CAENHVAsyn
CPP object's various createParamX
methods with X being replaced by Binary, Integer etc. This process of loading records is initiated by the CAENHVAsynConfig
call in the st.cmd.
The CAENHVAsyn
object acts as an adapter/facade to the objects it contains: crates, boards, channels. The CAENHVAsyn
object overrides asynPortDriver
methods, the overriding methods search through channel parameters, board parameters and system properties to find the value to set or get corresponding to the asyn call reason. These various methods often then make calls to the CAENHVWrapper
.
- Access the
EPICS_win10_x64
build from the instrument and deploy it to the instrument using the install_to_inst.bat script - When finished copying run
create_icp_binaries.bat
- Then copy
msvcr100.dll
(there are a few located around the install but not in the right place) toEPICS/crtl