Configure Mini Inst
Wiki > Deployment > Deployment on an Instrument Control PC > Configure Mini Inst
To install a mini inst, you need to install the EPICS backend and add the following in the \NDXxxxxx\configurations\
area.
-
startup.txt
: contains a list of the IOCs to start (use the IOC name as in the IOCs list) -
globals.txt
: add macro values you want the IOCs instartup.txt
to start up with. To do this, add a line per macro of the format<IOC_name>__<macro_name>=<macro_value>
You can start up the mini inst using the normal start_ibex_server
and stop it with stop_ibex_server
. It will look for the startup.txt
and start up the mini inst instead of the full IBEX server if it exists.
If you update a configuration in globals.txt then don't forget to restart the IOC. This can be done either by running start_ibex_server
but it is more efficient to console to the IOC and restart it using ctrl+x
.
If you want to run an IOC in a mini-inst then:
- Add the device name to the "startup.txt" (
C:\Instrument\Settings\config\NDX<instrument>\configurations\startup.txt
)- e.g. JSCO4180_01
- Then set macros in "globals.txt" (
C:\Instrument\Settings\config\NDX<instrument>\configurations\globals.txt
)- e.g. JSCO4180_01__PORT=COM13
- Run
make iocstartups
in the epics terminal. This adds the device to the allowed running devices. - Start IBEX server as normal
start_ibex_server.bat
- If you are running the IOC to allow a Vi in SECI to communicate then configure that vi (usually the pv prefix)
- e.g. Run
C:\LabVIEW Modules\Drivers\Jasco PU-4180 HPLC Pump\Jasco HPLC Pump - System.llb
- Initialisation File change IOC Root toIN:CRISP:JSCO4180_01:
- e.g. Run
- Run the Vi
ENGIN-X will run a mini-inst for the instron stress rig that needs to use both ARINST
and ARACCESS
. These need to be added to startup.txt
and MySQL needs to be installed and running on the machine.