The Instron stress rig is a National Instruments GPIB device. It requires some special setup to get going:
- The Ethernet GPIB box requires a driver for the LabVIEW driver. This can be installed from
\olympic\Babylon5\Public\GHowells\NI GPIB
- Run the installer as an administrator, accept the defaults. It will unzip, then install. It takes a while to install.
- You then need to set up the ENET box, Gareth or Freddie can show you how to set this up
- LabVIEW vi located at :
C:\LabVIEW Modules\Instruments\ENGINX\Stress Rig\Stress Rig - System Functions.llb\Stress Rig - 100 kN Stress Rig.vi
- On running the vi, you will get some dialogs – just ok through them. The indicators should then be updating.
- If you can't get the labview to talk at all, the stress rig might need to be power cycled. @GDH-ISIS and @FreddieAkeroyd know how to do this. The basic instructions are below:
- Turn off the stress rig PC (white/beige pc located under the stress rig)
- Turn off the GPIB box (this looks similar to a router, located under the stress rig near to the stress rig PC)
- Wait a few seconds
- Turn on the stress rig PC
- Wait several minutes. The stress rig will go through a self-test phase. You need to wait for this process to be complete.
- If you need to calibrate the rig, the following sequence on the console should be used: "Load channel: setup" -> "cal" -> "cal" -> "auto" -> "go" (if you are unsure about this ask the scientists)
- Turn back on the GPIB box.
- After the GPIB box has been turned on for a short time (e.g. 1 minute) the driver (LabVIEW or IOC) should be ready to connect.
The stress rig driver uses the following DB files:
-
controls.db
- provides basic PVs such as stopping the rig, getting the status, changing control channels, etc -
controls_channel.db
- provides PVs that are common to each channel (Position, Stress, Strain), e.g. setpoints, readbacks, step times -
controls_channel_specific.db
- add on PVs for the above - used to add PVs for some channels that don't exist on others -
controls_waveform.db
- provides the PVs dealing with the waveform generator -
logging.db
- provides the PVs to do with logging to a file
The protocol is defined in C:\Instrument\Apps\EPICS\support\instron\master\instronSup
(see https://github.com/ISISComputingGroup/EPICS-instron/blob/master/instronSup/devinstron.proto)
- Every "write" command (commands starting with C or P) must be preceded by
P909,1
(switch to computer control mode) andC904,0
(disable watchdog). For convenience there is thesetControlModeCom
function in the protocol file which does these for you. - Every "write" command (commands starting with C or P) must be followed by
P909,0
(switch to front panel control mode) - We have had issues with PVs dropping off to zero while being read from the stress rig. To solve this, use a
:RAW
record which does a read, and monitor it usingCP
from another record. This fixes the issue of getting zeroes. - PVs in the stress rig don't scan by themselves typically, they are triggered from one of two read loops:
-
READ_SLOW
scans at 1 Second interval -
READ_VAR
reads at a rate that can be varied by the user - In the
READ_VAR
loop the.1 Second
option has been removed, this is too fast for the rig to handle
-
- If something works in LabVIEW but not in EPICS, or vice-versa, NI Input/Output trace (NI Spy) can be very useful to compare the traffic and spot any differences.