Litron Laser Timing Control is an extension of the DG645 IOC made for HIFI to use. It replaces the LabView functionality as described here. To enable the Litron Timing mode in the DG645 IOC, put 'LITRON' in the 'APPLICATION' macro.
By setting a delay and offset the following calculation is done.
dC_0 = Delay + Offset
in us
Under the following constraints:
If mode is 1:
if dC_0 + Delay + Offset > 39900 us
or Delay + Offset < 0 us
then:
Error State
If mode is 2:
if Delay + Offset > 39900 us
or dC_0 + Delay + Offset <= 0 us
then:
Error State
When the user tries to set the mode then the following is applied:
If IOC mode is 1 or 2 then:
Set device mode to be 1 or 2 respectively
else: # IOC mode is auto
if Delay > Offset then:
Set device mode to 1
else:
Set device mode to 2
Note that IOC mode is not the same as device mode. IOC mode refers to a value of auto
or 1
or 2
, the device mode refers to which configuration/setting slot the DG645 is in (0-9). IOC mode 1 and 2 refers to device setting slots 1 and 2.
- Unlike the DG645 IOC- when the user changes the device mode, we do not restart the IOC. We chose not to as we need to know at any point what the device modes is for error checking- if we restart the IOC then we lose track of what the device mode is.
- The original spec for Litron Laser Timing Control states that
dA_0
and dC_0` should be polled at 4hz. We chose to keep the original polling rate of the DG645 at 1hz.